Synchrotron SAXS
data from solutions of
SWAXS from AT DNA duplex (25 base pair)
in
10 mM MOPS, 150 mM NaCl, 50 µM EDTA, pH 7
were collected
on the
16-ID (LiX) beam line
at the National Synchrotron Light Source II (NSLS-II) storage ring
(Upton, NY, USA)
using a Pilatus 1M detector
at a wavelength of λ = 0.0819 nm
(I(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle).
at 25°C.
10 successive
0.500 second frames were collected.
The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted.
The low angle data collected at lower concentration were merged with the highest concentration high angle data to yield the final composite scattering curve.
Sample detector distance = UNKNOWN. Concentration max = UNKNOWN